5.
Install the high voltage cables. Cable drape brackets are included with the
hardware accessories bag. The cable drape brackets are installed above
and below the handle on the right side of the tubestand. Loosely drape the
HV cables with the collimator and stator cables within the brackets until the
vertical slide has been freed to move. The cables will route directly into the
right rear of the table base, above the lower tubestand rail.