InnoVet Select
™
HF APR
Service Manual
63
Note: As kVp is increased, the filament value number is decreased slightly
due to the space charge effect.
Note: We recommended recording the final calibrated FIL#TBL and MASTER
mA/MASTER kV values from the XRAY CAL screen in the tables below for
future reference.
Table 9-2: For Recording FILAMENT AMPS Calibration Values
SMALL FILAMENT
LARGE FILAMENT
50 mA
100 mA
150 mA
200 mA
250 mA
300 mA
40
kVp
50
kVp
70
kVp
90
kVp
110
kVp
125
kVp
Table 9-3: For Recording MASTER kV and MASTER mA Calibration Values
MASTER mA Small
Focal Spot
MASTER mA Large Focal
Spot
MASTER kV
Note that in APR edit mode, only 100 mA small and 300 mA large are selectable.