InnoVet Select
™
HF APR
Service Manual
59
9.3 Step-by-Step
Calibration
1.
Connect the test equipment.
Connect test equipment per section 9.1.1 above.
2.
Enter Calibration Mode and turn off mA feedback.
On the System Controller board set DIP1- 4 to OFF to disable mA feedback. Turn
the Operator Console on while holding down the BACK button. The CAL BOOTUP
screen will display as shown below.
Select XRAY CAL, and enter the password (upper left, lower left, lower right, upper
right outer-most arrow keys). The XRAY CAL screen will display as shown below.
Select PWR MOD CAL. The PWR MOD CAL screen will display as shown below.
SET DATE CAL BOOTUP
& TIME
EDIT APR MODE
RENAME
TECH
TUBE ROLL CAL
MEMORY FUNC
XRAY CAL
EXIT TO RAD
XRAY CAL
TUBE SETUP
AEC WALL SET
PWR MOD SET
PWR MOD CAL
FIL# TBL
AEC TBL SET
MORE
POWER MOD
kVp STEP:40 CALIBRATION
FIL AMPS:4.30
mA 2
mA:50 SM
MASTER mA
TIME(ms):50 TABLE TOP SAVE
PAGE