5.3 Component
Access
5.3.1 Access to Operator’s Console PCBA
Access to the Operator Console PCBA is achieved by removing the two 10-32
screws located on the top rear of the operators console as viewed from the front,
and loosening the two bottom 10-32 screws which are slotted. The back cover will
then lift straight up and off.
5.3.2 Access to Power Module
Access to the Power Module is achieved by removing the two screws from each
side of the cover, located near the front, and two screws from the back, located
near the bottom. Lift the cover up slightly and back from the Power Module frame.
All cables will exit the front.
5.3.3 Access to the High Voltage Transformer
Terminal connections P1, P2, and GND can be accessed by removing the two
screws at the terminal cover on top of the unit. Cathode is left, anode is right. The
mAs meter connection is on the left behind the cathode well.
CAUTION: There is no access to the interior of the High Voltage
Transformer. This is a sealed unit and opening it in the field
without guidance by Summit Technical Support will void the
factory warranty.