70
Safety functions
EN
11
Safety functions
Automatic venting system
The device is equipped with an automatic venting system.
When the insufflator detects an overpressure it automatically activates the vent-
ing system. The venting system releases gas from the abdomen or from the ves-
sel until the set nominal value has been reached again.
The function "Automatic venting system" can be activated/deactivated in the
Configuration menu.
CAUTION!
Please note that the automatic venting system is only active during the Veress
insufflation mode if "In Veress insufflation ON" has been set in the configuration
menu (only possible in Pediatric operating mode, see chapter 7 Using and Con-
trolling the PNEUMO SURE High Flow Insufflator in Pediatric Operating Mode,
page 34).
WARNING!
The venting rate of the automatic venting system is limited. Always monitor the
actual pressure when using additional insufflation sources.
Insufflation source
The manufacturer advises against using additional, non-pressure controlled in-
sufflation sources during minimally invasive surgical procedures.
The self-inflating property of lasers cooled with CO2 and argon beamers can lead
to values exceeding the recommended and safe pressure rating.
Exceeding nominal pressure
After 2 to 5 seconds (depending on the setting in the Configuration menu), the
display depicts
Venting system active
when the nominal pressure has been ex-
ceeded by 2-5 mm Hg.
Exceeding nominal pressure for more than
5 seconds
If the overpressure cannot be reduced by the automatic venting system within 5
seconds, the display depicts
Overpressure
followed by
Venting system active
. An
acoustic alarm is emitted.
Nominal pressure limit 30 mm Hg /
20 mm Hg
Once the nominal pressure limit has been reached/exceeded (High Flow/Bariat-
ric -> 30 mm Hg or Pediatric/Vessel Harvest -> 20 mm Hg), the display depicts the
message
Overpressure
. A warning signal is emitted at the same time.
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......