66
Configuration Menu (Overview)
EN
10.2.4 Setting the Flow Safety Limit*
Factory setting = Limit OFF
*selectable only in Pediatric operating mode and Vessel Harvest operating mode.
In the Configuration menu I, press the
to access the Configuration menu II.
In the configuration menu II, press the
Flow safety limit
function field to access
the setting. (Only selectable while in Vessel Harvest or Pediatric operating mode;
otherwise depicted in gray - if set, valid for both modes)
The display switches to the
Flow safety limit
selection menu.
For example, press the
Limit ON
function field.
You now have the opportunity to
• save the selected value by pressing the
(SAVE)
function field. The display de-
picts
(SAVED)
for 2 seconds. After saving you return automatically to the pre-
vious level,
• press the
function field to return to the previous menu level without saving,
• or press
(EXIT)
to return to the work screen without saving.
10.2.5 Setting the Warning signal: Occlusion
Factory Settings
In the Configuration menu I, press the
to access the Configuration menu II.
The display changes to the Configuration menu II.
In the configuration menu II, press the
Warning signal: Occlusion
function field
to access the selection menu.
The display switches to the
Warning signal: Occlusion
selection menu.
For example, press the
Signal OFF
function field.
You now have the opportunity to
1. save the selected value by pressing the
(SAVE)
function field. The display de-
picts
(SAVED)
for 2 seconds. After saving you return automatically to the pre-
vious level,
2. press the
function field to return to the previous menu level without sav-
ing,
3. or press
(EXIT)
to return to the work screen without saving.
10.3
Utility menu
Utility menu
While insufflation is stopped, press the
Actual
function field for 2 seconds to ac-
cess the Configuration menu. In the Configuration menu I, press the
function
field to access the Configuration menu II and there press the
function field
Operating mode
Factory setting
Bariatric
Signal ON
High Flow
Signal ON
Vessel Harvest
Signal OFF
Pediatric
Signal ON
(1)
(3)
(2)
(1)
(3)
(2)
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......