Using and Controlling the PNEUMO SURE High Flow Insufflator in Pediatric Operating Mode
39
EN
Testing the device
1. The device is switched off, no tube set is connected.
2. Make sure the gas supply is connected and open.
3. Use the ON/OFF switch
(1)
to turn the device on. The device conducts a de-
vice check.
4. Select Pediatric operating mode in the insufflation mode selection.
5. Connect an original insufflation tube set
(6)
to the device.
6. Attach the insufflation tube to the Veress cannula
(8)
.
7. Attach the stopcock (valve)
(7)
to the Veress cannula.
8. If the gas consumption display function field does depict a value, press the
(4)
function field to reset the display to
0
.
9. Select the nominal pressure 8 mm Hg
(2)
and the gas flow 2 l/min
(5)
.
10. Start insufflation: Press the
START
function field
(3)
.
11. Insufflate for approx. 30 s. The display status line depicts
Occlusion
after max.
4 seconds.
12. Stop insufflation: Press the
STOP
function field
(3)
.
WARNING!
If the actual gas consumption is higher than 0.4 l, there is a leak in the system. If
this happens, use steps 13 to 15 outlined below to locate the leak.
13. Repeat items 7 to 11 without Veress cannula and with closed tube end. The
previously connected Veress cannula has a leak if gas consumption is now be-
low 0.4 l.
14. Repeat items 8 to 12 without Veress cannula and without tube if another leak
becomes apparent. Close the end of the insufflation tube connection for this
test. If the gas consumption is then below 0.4 l, the previously used tube set
has a leak.
15. If another leak is detected, this leak is then directly within the device. Make
sure the device can no longer be operated until a qualified service technician
conducts the appropriate tests and repairs.
WARNING!
Never work with a leaky insufflation tube, accessory, and/or device. This can
lead to an incorrect measurement of the actual pressure values, which can cause
an uncontrolled pressure increase in the abdomen.
Testing gas heating functionality
The functionality of the gas heating is indicated by displaying the gas heating
Fig. 7-1
Layout for testing the device
(1)
ON/OFF switch
(2)
Nominal pressure display
(3)
START/STOP function field
(4)
Gas consumption display
(5)
Nominal flow display
(6)
Insufflation tube set
(7)
Stopcock (valve)
(8)
Veress cannula
(1) (2) (3)
(4)
(5)
(6)
(7)
(8)
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......