48
Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Bariatric Operating Mode
EN
vice check.
4. Select Bariatric operating mode in the insufflation mode selection.
5. Connect an original insufflation tube set
(6)
to the device.
6. Attach the insufflation tube to the Veress cannula
(8)
.
7. Attach the stopcock (valve)
(7)
to the Veress cannula.
8. If the gas consumption display function field does depict a value, press the
(4)
function field to reset the display to
0
.
9. Select the nominal pressure 15 mm Hg
(2)
and the gas flow 5 l/min
(5)
.
10. Start insufflation: Press the
START
function field
(3)
.
11. Insufflate for approx. 30 s. The display status line depicts
Occlusion
after max.
4 seconds.
12. Stop insufflation: Press the
STOP
function field
(3)
.
WARNING!
If the actual gas consumption is higher than 0.4 l, there is a leak in the system. If
this happens, use steps 13 to 15 outlined below to locate the leak.
13. Repeat items 7 to 11 without Veress cannula and with closed tube end. The
previously connected Veress cannula has a leak if gas consumption is now be-
low 0.4 l.
14. Repeat items 8 to 12 without Veress cannula and without tube if another leak
becomes apparent. Close the end of the insufflation tube connection for this
test. If the gas consumption is then below 0.4 l, the previously used tube set
has a leak.
15. If another leak is detected, this leak is then directly within the device. Make
sure the device can no longer be operated until a qualified service technician
conducts the appropriate tests and repairs.
WARNING!
Never work with a leaky insufflation tube, accessory, and/or device. This can
lead to an incorrect measurement of the actual pressure values, which can cause
an uncontrolled pressure increase in the abdomen.
Testing gas heating functionality
The functionality of the gas heating is indicated by displaying the gas heating
symbol on the display when using the "Heated tube set with RTP for Pneumo
Sure".
"Real-Time Pressure Sensing" functionality
(RTP)
1. Connect the corresponding tube set ("Heated with Real-Time Pressure Sens-
ing" or "High Flow with Real-Time Pressure Sensing").
2. Insufflation is initially always started intermittently. Availability of the "Real-
Time Pressure Sensing" functionality is checked automatically. If this is the
case, the device switches to a continuous mode. This is indicated by the cor-
responding symbol on the display.
8.5.1
Filling Tube System with CO2
At least 1 l of CO2 has to be let out from the system before every surgery while
the tube is connected and the end of the tube is open. This will expel any air with-
in the tube system and the device.
1. Start insufflation: Press the
START
function field.
Wait until the gas consumption display shows
1.0 Liter
.
2. Stop insufflation: Press the
STOP
function field.
3. Press the gas consumption display field to reset the gas consumption display
to
0.0 Liter
. This ensures the gas consumption is correctly displayed during
surgery.
The function test is complete. The device is tested and ready to be used for sur-
gery.
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......