Operating the Device - General
23
EN
5.6.2
Displaying/Selecting Insufflation Operating Mode
The device can be equipped with up to 4 procedure modes. Procedure options can
be released or uploaded in different combinations depending on the order or
software upgrade.
1. Control or selection of the insufflation modes can be done only while insuf-
flation is stopped. Press the function field for 2 s to change the insufflation
operating mode.
2. The display depicts an insufflation operating modes overview. Unavailable
operating modes are depicted in gray and cannot be selected. Press the re-
spective function field to select the corresponding desired operating mode
(e.g. Pediatric).
3. After pressing the function field for the insufflation operating mode, the dis-
play depicts the selected procedure (e.g. Pediatric operating mode). The dis-
played parameters correspond with the factory settings or the values set in
the Configuration menu (see chapter 10.1 Configuration menu I, page 60).
5.6.3
Setting the Nominal Pressure - General
Press the function fields
º
or
»
on the display below the text line
Set Pressure
to
set the nominal pressure.
Applies to all modes:
• The nominal pressure can be increased or decreased during insufflation or
while insufflation is stopped.
• With every touch of the pressure function field
º
or
»
the nominal pressure is
increased/reduced in steps of 1 mm Hg. When pressing the pressure function
field
º
or
»
longer than 1,5 seconds, scrolling is activated but only up to the re-
spective, application-dependent safety threshold. Once this limit has been
reached, the status line depicts the message
Safety limit
and the actual value
flashes. To allow settings above the safety threshold release the pressure func-
tion field for 2 seconds. The display switches back to the status display. Now it
is possible to increase the pressure above the safety threshold but only to the
next safety threshold (as far as existent) or to the maximum setting.
• The latest working settings are stored with interruption of the insufflation, but
only if the last value was below the lowest safety threshold (
15 mm Hg for High
Flow and Bariatric operating mode, 12 mm Hg with Pediatric operating mode
and Vessel Harvest operating mode)
. Otherwise the set pressure value is reset
to the lowest safety threshold upon deactivation of insufflation.
• The latest settings are reset to the values preselected in the
Configuration
menu
when deactivating the device or when changing into the
Configuration
menu
or the Mode Selection menu. Upon choice of the respective working
mode the nominal settings as stored in the Configuration menu are displayed
(see chapter 10.1 Configuration menu I, page 60).
5.6.4
Setting the Nominal Flow - General
Press the function fields
º
or
»
on the display below the text line
Set Flow
to set
the nominal flow.
Applies to all modes:
PNEUMO SURE
High Flow operating mode / Pediatric operating mode
PNEUMO SURE XL
High Flow operating mode / Pediatric operating mode /
Bariatric operating mode / Vessel Harvest operating
mode
START
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Mode: High Flow
3
15
Change
Mode
Actual
l/min
0
Liter
0.0
0
GAS
(1)
(2)
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......