![Stryker PneumoSure Скачать руководство пользователя страница 52](http://html.mh-extra.com/html/stryker/pneumosure/pneumosure_instructions-for-use-manual_1385986052.webp)
46
Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Bariatric Operating Mode
EN
8.3
Presetting Nominal Pressure in Bariatric Operating Mode
The nominal pressure can be set during insufflation or while insufflation is
stopped. Values may range from
1 to max. 30 mm Hg
or the value set in the Con-
figuration menu.
Increasing/decreasing nominal pressure:
Briefly press the
º
or
»
function field to increase or decrease the pressure. Keep-
ing the
º
or
»
function field depressed longer than 1.5 seconds activates scroll-
ing in increments of 1.
Safety limit:
When
increasing
the nominal pressure, the status line of the display depicts
Safe-
ty limit
starting at
15 mm Hg
and the nominal value flashes. The nominal pres-
sure
15 mm Hg
is a threshold value. This is where the recommended range for
the intra-abdominal pressure ends. Pressing the nominal pressure
º
function
field again does not increase the pressure any further. Release the function field
at that point. The display switches back to the nominal value setting after 2 sec-
onds. Now you can set a value up to 30 mm Hg.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
8.4
Presetting Nominal Flow in Bariatric Operating Mode
The nominal flow can be increased or decreased during insufflation or while in-
sufflation is stopped.
1. Briefly press the
º
or
»
function field to increase or decrease the flow. The
nominal flow can be increased from
1 to max. 45 l/min
.
• Briefly press the corresponding field to set values in increments of 1.
• Keeping the
º
or
»
function field pressed longer than 1.5 seconds activates
scrolling through the gas flow rates 5, 25, 45 l/min or 45, 25, 5 l/min.
2. The preset values of the Configuration menu can be changed individually (see
chapter 10.1 Configuration menu I, page 60). Select a nominal gas flow be-
tween 1-45 l/min. The preset value is indicated in the nominal gas flow dis-
play. The values for the nominal gas flow refer to a device without connected
tube, filter, or instrument. Tube, filter, and instrument can reduce the gas
flow.
The device monitors the gas flow in two different operating modes:
•
Veress insufflation (1-5 l/min)
•
Insufflation: Bariatric (6-45 l/min)
Veress insufflation operating mode:
Veress insufflation
is a gentle type of insufflation so that the actual pressure
does not exceed the preset pressure even in case of small volumes. To minimize
the risks in case of a faulty incision, the manufacturer recommends using
Veress
insufflation
to start a procedure (filling abdomen with CO2).
WARNING!
Please note that the automatic venting system is not active during Veress insuf-
flation mode.
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......