Using and Controlling the PNEUMO SURE High Flow Insufflator in Pediatric Operating Mode
37
EN
the intra-abdominal pressure ends. Pressing the nominal pressure ºfunction field
again does not increase the pressure any further. The status line of the display
depicts
Safety limit
and the nominal value flashes.
Release the function field at that point. The display switches back to the nominal
value setting after 2 seconds. Now you can set a value up to
20 mm Hg
.
7.4
Presetting Nominal Flow in Pediatric Operating Mode
The nominal flow can be set during insufflation or while insufflation is stopped.
Reducing nominal gas flow:
Press the nominal gas flow function field
»
to reduce the nominal gas flow. The
nominal gas flow is decreased in the range of
• 0.1 l/min to 2 l/min in increments of 0.1 l/min.
• 2 l/min to 20 l/min in increments of 1 l/min.
Keeping the
»
function field depressed longer than 1.5 seconds activates scroll-
ing.
• In the range of 0.1...2 l/min, rounding down to 1 or 0.1 l/min is applied.
• In the range of 2...20 l/min in increments of 1.
The nominal flow can be set during insufflation or while insufflation is stopped.
Increasing nominal gas flow:
Press the nominal gas flow function field
º
to increase the nominal gas flow. The
nominal gas flow is increased in the range of
• 0.1 l/min to 2 l/min in increments of 0.1 l/min.
• 2 l/min to 20 l/min in increments of 1 l/min.
Keeping the
º
function field depressed longer than 1.5 seconds activates scroll-
ing.
• In the range of 0.1 l/min...<2 l/min, rounding up to 1 or 2 l/min is applied.
• Then in one-step increments down to 20 l/min.
Safety limit:
When increasing the nominal flow, the status line of the display depicts
Safety
limit
starting at
5 l/min
and the nominal value flashes. (This safety limit can be
activated or deactivated in the Configuration menu).
Release the function field to increase the nominal gas flow further. The display
switches back to the nominal value setting after 2 seconds. Now you can set a
value of up to 20 l/min.
WARNING!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
START
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Safety limit
0.1
15
0
Liter
283
0
Actual
l/min
Change
Mode
GAS
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Safety limit
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......