Operating the Device - General
17
EN
5.3
Touch screen display
The above depiction of the touch screen also shows all display and function
fields. Field
(9)
serves as actual flow display (depicted without frame while insuf-
flating) and also as menu function field (depicted with frame).
Field
(12)
serves as insufflation operating mode display (depicted without frame
while insufflating) and also as control field for selecting the insufflation operat-
ing mode (depicted with frame).
Press the function field
(9)
or
(12)
depicted with frame and hold for 2 seconds to
trigger functions or set values. Additional explanations for individual elements
are presented in the subsequent respective control element descriptions.
Gas supply displays
The status of the gas supply is monitored by the device and indicated with sym-
bols and acoustic signals (see chapter 11 Safety functions for gas pressure display
information).
Gas supply with gas bottle
The following gas bottle pressures are displayed:
If gas supply pressure declines further, there are warnings to remind the user to
replace the gas tank immediately. Five warning signals can be heard and the
message "Check gas supply" is displayed at < 5 bar and again at 0 bar. Insufflation
stops at 0 bar.
Fig. 5-3
Screen displays
(1)
Continuous pressure reading dis-
play
(2)
Gas supply display
(3)
Actual pressure display
(4)
Gas consumption display/func-
tion field for reset
(5)
Gas heating connected/ready
(6)
Increasing nominal gas flow
(7)
Nominal gas flow display
(8)
Decreasing nominal gas flow
(9)
Actual gas flow display/ menu
function field
(10)
START/STOP function field
(11)
Status display/error and warning
messages
(12)
Insufflation operating mode dis-
play/selecting insufflation operat-
ing mode
(13)
Decreasing nominal pressure
(14)
Nominal pressure display
(15)
Increasing nominal pressure
Liter
0.0
START
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Change
Mode
Actual
Mode: High Flow
3
15
0
l/min
0
GAS
HEA
TING
(4)
(2)
(3)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
(1)
> 50 bar
40 - 50 bar
30 - 40 bar
15 - 30 bar; Three warning signals can be heard and the message
"Change gas tank" is displayed. User is advised to obtain a replace-
ment tank.
< 15 bar; Three warning signals can be heard and the message
"Check gas supply" is displayed. Replace gas tank immediately.
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......