Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Vessel Harvest Operating Mode
53
EN
9.3
Presetting Nominal Pressure in Vessel Harvest Operating Mode
The nominal pressure can be set during insufflation or while insufflation is
stopped. Values may range from
1 to max. 20 mm Hg
or the value set in the Con-
figuration menu.
Increasing/decreasing nominal pressure:
Briefly press the
º
or
»
function field to increase or decrease the pressure. Keep-
ing the
º
or
»
function field depressed longer than 1.5 seconds activates scrolling
in increments of 1.
Safety limit:
When
increasing
the nominal pressure, the status line of the display depicts
Safe-
ty limit
starting at
12 mm Hg
and the nominal value flashes.
The nominal pressure value
12 mm Hg
is a limit value and should not be exceed-
ed when performing vessel harvesting surgery. Pressing the nominal pressure
º
function field again does not increase the pressure any further.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
Release the function field at that point. The display switches back to the nominal
value setting after 2 seconds. Now you can set a value up to
15 mm Hg
.
The
Safety limit
message is re-displayed in the status line starting at 15 mm Hg
and the nominal value flashes.
The nominal pressure value
15 mm Hg
is a limit value and should not be exceed-
ed when performing vessel harvesting surgery. Pressing the nominal pressure
º
function field again does not increase the pressure any further.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
Release the function field at that point. The display switches back to the nominal
value setting after 2 seconds. Now you can set a value up to
20 mm Hg
.
9.4
Presetting Nominal Flow in Vessel Harvest Operating Mode
The nominal flow can be set during insufflation or while insufflation is stopped.
Increasing nominal gas flow:
• Press the function field
º
to increase gas flow. The nominal flow can be in-
creased from
1 to max. 10 l/min
.
• Briefly press the corresponding function field to set values in increments of 0.5
for a range from 1 to 5 l/min.
• Set values in increments of 1 for a range from 5 to 10 l/min.
• Keeping the
º
or
»
function field pressed longer than 1.5 seconds activates
scrolling through the gas flow rates 1, 4, 10 l/min (in the Configuration menu).
Reducing nominal gas flow:
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......