Operating the Device - General
EN
Time Pressure Sensing" functionality is checked automatically. If this is the case,
the device switches to a continuous mode. This is indicated by the corresponding
symbol on the display:
The
symbol is removed if continuous pressure measurement is currently not
possible or if insufflation is stopped.
CAUTION!
A closed, obstructed or disconnected pressure sensing line will disable the real-
time pressure sensing function. In this case the device will operate in the conven-
tional intermittent insufflation mode.
Error detection and monitoring of the "Real-
Time Pressure Sensing (RTP)" function
The continuous pressure measurement function is verified during the initial de-
vice self check. Should a defect within the measuring system be detected, three
acoustic warning signals are emitted. The symbol
is crossed out and the sta-
tus line reads as follows:
RTP defective / Call Service
.
A detected defect within continuous pressure measurement does not generally
hinder use of the device, however without the RTP function. Disconnect the mea-
suring channel and perform the surgery in the intermittent insufflation mode.
The acoustic warning on the defective RTP will be emitted with every activation/
deactivation of the insufflation as long as the pressure measurement function
remains defective.
CAUTION!
Do not attach a trocar to the pressure measurement / sensor line if the message
"RTP defective / Call Service" is displayed. Perform surgery only in the conven-
tional intermittent insufflation mode without using the real-time pressure sens-
ing function.
In case of a closed sensor line or with pressure on the sensor or the insufflation
line during the initial device self check the real-time pressure sensing function
will not be activated and the
will be crossed out. Three acoustic warning sig-
nals are emitted, the status line readsas follows:
RTP deactivated
. The device can
be operated in the conventional intermittent insufflation mode. In order to acti-
vate the RTP function the sensor line must be cleared, possible pressure released
and the device rebooted by turning it off and back on.
CAUTION!
When working with a tube set permitting the use of the RTP function, please
make sure that both lines are open upon activation of the device and there is
pressure neither on the sensor line nor on the insufflation line. Otherwise the
real-time pressure sensing function (RTP) will not be recognized and activated
during the initial device self check.
In case of an occlusion or a leakage in the sensor line during insufflation the de-
vice will automatically switch from the RTP operating mode into the convention-
al intermittent insufflation mode.
Leakage detection with function "Real-
Time Pressure Sensing (RTP)"
If the system detects a leak at the RTP connection regardless of the insufflation
mode:
• 2 long acoustic warning signal are emitted,
• the RTP symbol is no longer depicted,
• the message
Check/change tube set
is displayed,
• the usual intermittent pressure measuring is enabled again,
• the function of the occlusion detection is disabled.
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......