Gas heating ON/OFF
1.
The gas heating is automatically started at the START of the insufflation:
Press the
START
function field. The gas is automatically heated to 37 °C.
2.
Stopping gas heating:
Press the
STOP
function field. Gas heating is switched off.
3. Pull the tube set plug from the device connection.
Incorrect/defective heating tube
When a defective heating tube set is used,
• the display depicts a crossed out symbol
• the text line reads
Gas heater defective <-> Call service
,
• and an acoustic signal is emitted (3 short beeps).
Insufflation can be continued in this case, but the heating function is not avail-
able.
Gas temperature exceeds 42 °C
The device is equipped with a temperature sensor to protect against overheating,
caused for exemple by external heat sources. If the temperature sensor mea-
sures a temperature >42°C, the status field of the display depicts
Gas tempera-
ture >42 °C
; 3 accoustic signals are emitted in addition, the
HEATING
symbol is
crossed out
. Insufflation and the heating function are halted. After 3 seconds,
the display depicts
Disconnect luer lock
alternating with
Cool down tube
until
the insufflation is manually restarted.
WARNING!
Unplug the cable of the heating tube from the device if the temperature sensor
measures a gas temperature exceeding >42 °C. Hot gas in the abdomen can lead
to serious injuries.
1. Disconnect the insufflation tube from the trocar or Veress needle.
2. Press the
START
function field. The device insufflates without heating the
gas.
3. Let hot gas escape until the tube is only warm to the touch and then recon-
nect the device again.
4. Continue surgery without gas heating.
Check gas heating after surgery
1. Check gas heating after surgery using a different tube. Turn the device off
and back on after approx. 10 seconds have expired. Gas heating is reactivat-
ed.
2. Should the error message be displayed again, you can continue using the de-
vice without gas heating by observing the risks for hypothermia.
3. Call an authorized service technician to check/fix the gas heating.
5.6.1
Using the direct pressure measurement function (Real-Time
Pressure Sensing RTP)
The insufflation tube sets "Heated tube set with Real-Time Pressure Sensing
(RTP)" and "High Flow tube set with Real-Time Pressure Sensing (RTP)" are
equipped with a sensor line which enables continuous measuring of the abdom-
inal pressure. The insufflation line must be connected to one trocar and the mea-
suring line (sensor line) to a second trocar. This set-up enables the PNEUMO SURE
XL High Flow Insufflator to directly measure the actual abdominal pressure while
in
High Flow-
,
Pediatric-
and
Bariatric
operating mode. The "Real-Time Pressure
Sensing" is deactivated during the
Vessel Harvest
operating mode. The use of an
insufflation tube without a second measurement/sensor line only allows the in-
termittent measuring of the pressure.
Insufflation is initially always started intermittently. Availability of the "Real-
HEA
TING
START
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Liter
0.0
3
15
0
Change
Mode
Actual
Gas heater defective
l/min
0
GAS
HEA
TING
HEA
TING
0
Liter
0.0
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
3
15
Cool down tube
Change
Mode
Actual
l/min
0
START
GAS
HEA
TING
Pull apart at seam as
much as needed
at second trocar
at insufflation trocar
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......