Safety functions
73
EN
NOTE!
Please make sure that neither of the two tubes is connected to a trocar or closed
off during a device check. The continuous pressure measurement cannot be acti-
vated otherwise.
Power failure
If the power fails for less than 1 second, all settings are retained. If the power fails
for an extended time period, the device will function as it does when it is being
restarted.
If a device malfunction occurs that prohibits any further use of the device or di-
minished its safety, the display depicts only
Text message - >Call service
The following text messages may be displayed:
•
Contamination
•
Electronic defective
•
Sensor defective
•
Venting valve defective
•
Calibration error
•
Temperature error
Make sure the device can no longer be operated until a qualified service techni-
cian conducts the appropriate tests and repairs.