Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Vessel Harvest Operating Mode
55
EN
vice check.
4. Select Vessel Harvest operating mode in the insufflation mode selection.
5. Connect an original insufflation tube set
High Flow 2
(6)
to the device.
6. Press the function field for the gas consumption display
(4)
to set the display
to
0 Liter
.
7. Select the nominal pressure 10 mm Hg
(2)
and the gas flow 1 l/min
(5)
.
8. Close the end of the tube (Luer Lock)
(7)
.
9. Start insufflation: Press the
START
function field
(3)
.
10. Insufflate for approx. 30 s. The display status line depicts
Occlusion
after max.
4 seconds.
11. Stop insufflation: Press the
STOP
function field
(3)
.
WARNING!
If the actual gas consumption is higher than 0.4 l, there is a leak in the system.
12. Replace the tube set and repeat the test from step 6 to step 11.
13. If another leak is detected, this leak is then directly within the device. Make
sure the device can no longer be operated until a qualified service technician
conducts the appropriate tests and repairs.
WARNING!
Never work with a leaky insufflation tube, accessory, and/or device. This can
lead to an incorrect measurement of the actual pressure values, which can cause
an uncontrolled pressure increase in the abdomen.
Testing gas heating functionality
The functionality of the gas heating is indicated by displaying the gas heating
symbol on the display when using the "Heated tube set with RTP for Pneumo
Sure".
9.5.1
Filling Tube System with CO2
At least 1 l of CO2 has to be let out from the system before every surgery while
the tube is connected and the end of the tube is open. This will expel any air with-
in the tube system and the device.
1. Start insufflation: Press the
START
function field.
Wait until the gas consumption display shows
1.0 Liter
.
2. Stop insufflation: Press the
STOP
function field.
3. Press the gas consumption display field to reset the gas consumption display
to
0.0 Liter
. This ensures the gas consumption is correctly displayed during
surgery.
The function test is complete. The device is tested and ready to be used for sur-
gery.
WARNING!
Do not use this device if a defect is suspected or detected during the function
check. This also applies to obvious defects, especially defects on the power plug
and power cable.
9.6
Using the Device during Surgery
Using the device during surgery
WARNING!
The function test must be performed prior to each surgery.
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......