![Stryker PneumoSure Скачать руководство пользователя страница 51](http://html.mh-extra.com/html/stryker/pneumosure/pneumosure_instructions-for-use-manual_1385986051.webp)
Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Bariatric Operating Mode
45
EN
WARNING!
Embolism
Improper placement of the insufflation instrument could cause insufflation of
gas into a vessel, resulting in air or CO
2
embolisms. To reduce the risk of air or
CO
2
embolism, perform initial insufflation at a low flow rate and ensure that the
insufflation instrument is correctly positioned. Check the position of the insuf-
flation instrument immediately if the actual pressure rapidly reaches the nomi-
nal pressure value. CO
2
embolisms can also be caused by a high intra-abdominal
pressure. Avoid high-pressure settings and close damaged blood vessels at once.
WARNING!
Additional insufflation sources
The use of additional insufflation sources increases the intra-abdominal pres-
sure. Continuously monitor intra-abdominal pressure over the course of the en-
tire insufflation if additional sources are used.
WARNING!
Device-inherent dangers
Please read the general risks and dangers information in chapter 2.1 Hazards,
page 5 and chapter 3.3 General Device-Inherent Dangers, page 8.
8.2
Selecting Bariatric Operating Mode
1. Selection of the insufflation mode can be done only while insufflation is
stopped. If Bariatric operating mode is not yet set and displayed, press the
function field
Change Mode
for 2 seconds to get the insufflation mode over-
view.
2. The display depicts an insufflation modes overview. Select the desired insuf-
flation operating mode by pressing the
Bariatric
function field. *Inactive op-
erating modes are displayed in gray.
3. The procedure profile is depicted on the display. The displayed parameters
correspond with the factory settings or the values set in the Configuration
menu (see chapter 10.1 Configuration menu I, page 60).
START
Set Pressure
Actual Pressure
mm Hg
mm Hg
Set Flow
l/min
Mode: High Flow
3
15
Change
Mode
Actual
l/min
0
Liter
0.0
0
GAS
(1)
(2)
Содержание PneumoSure
Страница 6: ......
Страница 101: ...Appendix 95 EN 19 Appendix 19 1 Test Log Date Results Comment Signature ...
Страница 104: ......
Страница 107: ...3 ES 19 Apéndice 100 19 1 Registro de test 100 Índice alfabético 101 ...
Страница 108: ......
Страница 204: ...100 Apéndice ES 19 Apéndice 19 1 Registro de test Fecha Resultado Anotaciones Firma ...
Страница 209: ...3 FR 19 Annexe 99 19 1 Procès verbal de tests 99 Index alphabétique 100 ...
Страница 210: ......
Страница 305: ...Annexe 99 FR 19 Annexe 19 1 Procès verbal de tests Date Résultat Remarque Signature ...
Страница 308: ......
Страница 405: ...Anexo 97 PT 19 Anexo 19 1 Protocolo de teste Data Resultado Anotações Assinatura ...
Страница 408: ......
Страница 409: ......