Table 9 – Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communications equipment
(according to IEC 60601-1-2)
Test frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
Power
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 – 390
TETRA 400
Pulse modulation
18 Hz
1.8
0.3
27
450
430 – 470
GMRS 460,
FRS 460
FM
± 5 kHz deviation
1 kHz sine
2
0.3
28
710
704 – 787
LTE Band 13, 17
Pulse modulation
217 Hz
0.2
0.3
9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
18 Hz
2
0.3
28
870
930
1 720
1 700 – 1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse modulation
217 Hz
2
0.3
28
1 845
1 970
2 450
2 400 – 2 570
Bluetooth,
WLAN 802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse modulation
217 Hz
2
0.3
28
5 240
5 100 – 5 800
WLAN 802.11 a/n
Pulse modulation
217 Hz
0.2
0.3
9
5 500
5 785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ENDOMAT
®
SELECT model UP 210
may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.