Table 5 – Input a.c. power PORT
(according to IEC 60601-1-2)
Phenomenon
Basic EMC standard
IMMUNITY TEST LEVELS
Professional healthcare
facility environment
HOME HEALTHCARE
ENVIRONMENT
Electrical fast transients/bursts
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
±0.5 kV, ±1 kV
Surges
Line-to-ground
IEC 61000-4-5
±0.5 kV, ±1 kV, ±2 kV
Conducted disturbances
induced by RF fields
IEC 61000-4-6
3 V
0.15 MHz – 80 MHz
6 V in ISM bands
between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
Not applicable since released only
for professional healthcare facility
environment
Voltage dips
IEC 61000-4-11
0 %
U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 %
U
T
; 1 cycle
and
70 %
U
T
; 25/30 cycles
Single phase: at 0° Grad
Voltage interruptions
IEC 61000-4-11
0 %
U
T
; 250/300 cycles