74
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Electromagnetic
Compatibility
(EMC) Information
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Guidance and manufacturer’s declaration – electromagnetic emissions
ENDOMAT
®
SELECT model UP 210 is intended for use in the electromagnetic environment specified below.
The customer or user of the ENDOMAT
®
SELECT model UP 210 should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
acc. to CISPR 11
Group 1
ENDOMAT
®
SELECT model UP 210 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause interference in
nearby electronic equipment.
RF emissions
acc. to CISPR 11
Class B
The ENDOMAT
®
SELECT model UP 210 is suitable for use in all establishments
including domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic purposes.
–
Professional healthcare facility environment
–
Harmonic emissions
acc. to IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
acc. to IEC 61000-3-3
Complies
Accessories and cables which have been shown to comply with IEC 60601-1-2:
Type
Shielded
Length [m]
Ferrite
Used for
PA
No
> 3
No
Potential equalization
Power cord
No
3
No
Power supply