75
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Electromagnetic
Compatibility
(EMC) Information
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Table 4 – ENCLOSURE PORT
(according to IEC 60601-1-2)
Phenomenon
Basic EMC standard or
test method
IMMUNITY TEST LEVELS
Professional healthcare
facility environment
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC DISCHARGE
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM fields
IEC 61000-4-3
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Not applicable since released only
for professional healthcare facility
environment
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
See table 9
RATED power frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz