56
Technische Beschreibung
Technical description
Descripciones técnicas
Information messages
Brief description (1 line with approx. 40 characters possible)
Response
Possible cause
0x768: Direction change
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the speedometer
0x770: Defective electronics (PS)
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Error in the monitoring system
0x771: SW error (PS)
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Error in the monitoring system
0x772: PS error (PUST)
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Error in the monitoring system
0x773: PS error (ST)
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Error in the monitoring system
0x774: PS error (H off)
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Error in the monitoring system
0x775: PS error (H on)
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Error in the monitoring system
0x780: Voltage error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Operating voltage outside of permissible range
0x790: WD task error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
BE: Program crash