Technische Beschreibung
Technical description
Descripciones técnicas
Information messages
Brief description (1 line with approx. 40 characters possible)
Response
Possible cause
300: High pressure
Observe surgical
fi
eld, ensure pressure reduction
Stopcock on instrument closed
301: Maximum pressure
Observe surgical
fi
eld, ensure pressure reduction
Excess pressure alarm in URO and HYS application if 150 mmHg is exceeded
500: Main functions not active
Acoustic and visual signal – device can still be operated:
A touchscreen calibration can be started by pressing the touchscreen for 25 seconds. This
text message will appear after 25 seconds.
Remedy: Check touchscreen and clean if necessary. This text message may also appear in
the event of a touchscreen short circuit. In this case: Turn the device off and on again, and
contact KARL STORZ Service if the error occurs again.
FE: A touch screen activation lasting more than 25 s has been detected
501: Touch calibration
This text message is part of the touchscreen calibration procedure (see text message 500)
and is displayed after the touchscreen is touched continuously for 30 seconds.
The screen must be released and touched again within 5 seconds in order to start
calibration.
FE: A touch screen activation lasting more than 30 s has been detected
0x750: Defective electronics
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the cartridge module
0x760: Defective drive
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the drive control unit
0x761: Defective control unit
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the general control unit
0x762: Defective H bridge
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the H bridge
0x763: Motor current offset
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the motor current measurement
0x764: Peak current
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error, overcurrent
0x765: Index missing
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the speedometer
0x766: Index moved
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Electronics error in the speedometer
0x767: Motor PWM
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Control error in the motor