Filament Replacement 8-13
SRS Residual Gas Analyzer
15. Visually inspect the filament alignment and do any adjustments that might be
necessary. The filament should form a circle around the anode grid. Slight bends in
the filament are common, and do not compromise its performance. Severe bends
might result in electrical-shorts to the repeller and anode grid and need to be
corrected. Use gentle pressure on the filament wire to bend it back into its correct
shape if needed (Note: use a clean cotton swab for this procedure).
16. Attach the repeller to the longer filament rod using a fresh screw. Align the cage and
tighten the screw (Correct alignment is best assured when the two small holes on the
side of the repeller cage line up with the filament screws.)
17. Inspect visually the entire ionizer assembly to assure its correct alignment, and, if
satisfied, mount the probe back on the vacuum system. Check the correct rotational
orientation of the probe (i.e. same as it was before disassembly) before tightening the
flange bolts.
18. A bakeout of the probe is recommended before operating the RGA again.
Содержание RGA100
Страница 4: ...SRS Residual Gas Analyzer iv...
Страница 18: ...xviii Command List SRS Residual Gas Analyzer...
Страница 46: ...2 14 Residual Gas Analysis Basics SRS Residual Gas Analyzer...
Страница 66: ......
Страница 78: ...4 12 Mass Filter Power supply SRS Residual Gas Analyzer...
Страница 104: ......
Страница 107: ...Programming the RGA Head 6 3 SRS Residual Gas Analyzer Error Byte Definitions 6 69...
Страница 216: ...8 26 Quadrupole filter cleaning SRS Residual Gas Analyzer...
Страница 246: ...11 2 SRS Residual Gas Analyzer...
Страница 247: ......
Страница 248: ......
Страница 268: ...Appendix B SRS Residual Gas Analyzer 7...
Страница 310: ...Appendix D SRS Residual Gas Analyzer 25 Select TCP IP and click the Properties button to find the IP address...
Страница 312: ...Appendix D SRS Residual Gas Analyzer 27...
Страница 313: ...28 Appendix D SRS Residual Gas Analyzer...