Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting options
Access
level
10064
Float / 4
Upper range
value ROC
Upper rate of change
value that is mapped to
the upper nominal
range boundary selec‐
ted by Current mode
(8806). Applicable if
Process value (8802) =
36.
1.0
[m/s²]
(8564)
-
Read /
write
10066
Float / 4
Lower range
value ROC
Lower rate of change
value that is mapped to
the lower nominal
range boundary selec‐
ted by current mode
(8806). Applicable if
Process value (8802) =
36.
0.0
[m/s²]
(8564)
-
Read /
write
8807
Float / 4
Damping value
Time constant of damp‐
ing filter for current out‐
put signal.
0.0
[s]
0.0 - 100.0
Read /
write
8814
Unsigned / 2
Fail-safe behav‐
ior
Current output reaction
in case of a fault.
0
● 0: Lower fault current
● 1: Upper fault current
● 2: Last valid value
● 3: Disabled
● 4: Custom value
Read /
write
8815
Float / 4
Fail-safe value
Output value in case of
a fault and when Fail-
safe behavior (8814) is
configured to 4.
0.0
[mA]
0.0 - 25.0
Read /
write
7146
Float / 4
Forced value
Forced value when
forcing is active.
4.0
[mA]
3.5 - 25.0
Read /
write
8886
Float / 4
Offset adjust‐
ment
Command to trim the
lower range value of
the loop current.
Simulate a loop current
of 4.0 mA, measure the
current with an ampere
meter and write the
measured value to this
register.
[mA]
0.0 - 25.0
Write on‐
ly
8888
Float / 4
Gain adjustment Command to trim the
upper range value of
the loop current.
Simulate a loop current
of 20.0 mA, measure
the current with an am‐
pere meter and write
the measured value to
this register.
[mA]
0.0 - 25.0
Write on‐
ly
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
85
Содержание SITRANS F Coriolis FCT030
Страница 6: ...Table of contents FCT030 Modbus From firmware 4 0 6 Function Manual 06 2018 A5E39931813 AB ...
Страница 16: ...Communication 2 4 Connecting FCT030 Modbus From firmware 4 0 16 Function Manual 06 2018 A5E39931813 AB ...
Страница 194: ...Definitions A 1 Definitions FCT030 Modbus From firmware 4 0 194 Function Manual 06 2018 A5E39931813 AB ...
Страница 198: ...Index FCT030 Modbus From firmware 4 0 198 Function Manual 06 2018 A5E39931813 AB ...