Modbus
register
Data type /
Size in bytes
Parameter
Description Default value
[units]
(units register)
Value range /
Setting options
Access
level
7609
Unsigned / 2
Density status
Quality sta‐
tus code of
the process
value Den‐
sity
● Limit status bit 4 & 5:
● 11: Constant
● 01: Low limited
● 10: High limited
● 00: Not limited
● Process data status bit 6 & 7:
● 11: Good
● 01: Poor Accuracy
● 10: Manual / Fixed
● 00: Bad
7709
Unsigned / 2
Media tempera‐
ture status
Quality sta‐
tus code of
the process
value Me‐
dia temper‐
ature
● Limit status bit 4 & 5:
● 11: Constant
● 01: Low limited
● 10: High limited
● 00: Not limited
● Process data status bit 6 & 7:
● 11: Good
● 01: Poor Accuracy
● 10: Manual / Fixed
● 00: Bad
7714
Unsigned / 2
Frame tempera‐
ture status
Quality sta‐
tus code of
the process
value
Frame tem‐
perature
● Limit status bit 4 & 5:
● 11: Constant
● 01: Low limited
● 10: High limited
● 00: Not limited
● Process data status bit 6 & 7:
● 11: Good
● 01: Poor Accuracy
● 10: Manual / Fixed
● 00: Bad
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
65
Содержание SITRANS F Coriolis FCT030
Страница 6: ...Table of contents FCT030 Modbus From firmware 4 0 6 Function Manual 06 2018 A5E39931813 AB ...
Страница 16: ...Communication 2 4 Connecting FCT030 Modbus From firmware 4 0 16 Function Manual 06 2018 A5E39931813 AB ...
Страница 194: ...Definitions A 1 Definitions FCT030 Modbus From firmware 4 0 194 Function Manual 06 2018 A5E39931813 AB ...