Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting options
Access
level
9423
Float / 4
Upper range
value tempera‐
ture
Upper medium temper‐
ature value that is map‐
ped to the upper nomi‐
nal range boundary se‐
lected by Frequency
value high (9435). Ap‐
plicable if Process val‐
ue (9404) = 3.
Max. medium
temperature
[°C]
(7700)
-
Read /
write
9425
Float / 4
Lower range
value tempera‐
ture
Lower medium temper‐
ature value that is map‐
ped to the lower nomi‐
nal range boundary se‐
lected by Frequency
value low (9437). Appli‐
cable if Process value
(9404) = 3.
Min. medium
temperature
[°C]
(7700)
-
Read /
write
9427
Float / 4
Upper range
value standard
volume flow
Upper standard volume
flow value that is map‐
ped to the upper nomi‐
nal range boundary se‐
lected by Frequency
value high (9435). Ap‐
plicable if Process val‐
ue (9404) = 27.
0.4 % • max.
standard vol‐
ume flow
[Sm
3
/h]
(7964)
-
Read /
write
9429
Float / 4
Lower range
value standard
volume flow
Lower standard volume
flow value that is map‐
ped to the lower nomi‐
nal range boundary se‐
lected by Frequency
value low (9437). Appli‐
cable if Process value
(9404) = 27.
-
[Sm
3
/h]
(7964)
-
Read /
write
10236
Float / 4
Upper range
value current
Upper current value
that is mapped to the
upper nominal range
boundary selected by
Frequency value high
(9435). Applicable if
Process value (9404) =
25 or 26.
20.0
[mA]
-
Read /
write
10238
Float / 4
Lower range
value current
Lower current value
that is mapped to the
lower nominal range
boundary selected by
Frequency value low
(9437). Applicable if
Process value (9404) =
25 or 26.
4.0
[mA]
-
Read /
write
9408
Float / 4
Damping filter
Time constant of damp‐
ing filter for frequency
output signal.
0.0
[s]
0.0 - 100.0
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
114
Function Manual, 06/2018, A5E39931813-AB
Содержание SITRANS F Coriolis FCT030
Страница 6: ...Table of contents FCT030 Modbus From firmware 4 0 6 Function Manual 06 2018 A5E39931813 AB ...
Страница 16: ...Communication 2 4 Connecting FCT030 Modbus From firmware 4 0 16 Function Manual 06 2018 A5E39931813 AB ...
Страница 194: ...Definitions A 1 Definitions FCT030 Modbus From firmware 4 0 194 Function Manual 06 2018 A5E39931813 AB ...
Страница 198: ...Index FCT030 Modbus From firmware 4 0 198 Function Manual 06 2018 A5E39931813 AB ...