Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
8780
Unsigned / 2
Recipe 2 unit
corrected vol‐
ume flow
Corrected volume flow
unit for dosing parame‐
ters
166
166: normal cubic meters
167: normal liters
168: standard cubic feet
171: standard liters
172: standard cubic me‐
ters
Read /
write
8708
Float / 4
Recipe 2
amount mass
flow
Target amount for dos‐
ing mass flow
-
-
Read /
write
8710
Float / 4
Recipe 2
amount volume
flow
Target amount for dos‐
ing volume flow
-
-
Read /
write
8712
Float / 4
Recipe 2
amount correc‐
ted volume flow
Target amount for dos‐
ing corrected volume
flow
-
-
Read /
write
7586
Unsigned / 2
Recipe 2 Dec.
point mass flow
Specify the selected
number of decimals for
mass flow
-
-
Read /
write
7587
Unsigned / 2
Recipe 2 Dec.
point volume
flow
Specify the selected
number of decimals for
volume flow
-
-
Read /
write
7588
Unsigned / 2
Recipe 2 Dec.
point corrected
volume flow
Specify the selected
number of decimals for
corrected volume flow
-
-
Read /
write
8715
Float / 4
Recipe 2 fixed
compensation
Mass flow
Fixed compensation
mass flow value that is
used to close the
valve(s) before the
measured amount val‐
ue has reached the tar‐
get amount
-
-
Read /
write
8717
Float / 4
Recipe 2 fixed
compensation
Volume flow
Fixed compensation
volume flow value that
is used to close the
valve(s) before the
measured amount val‐
ue has reached the tar‐
get amount
-
-
Read /
write
8719
Float / 4
Recipe 2 fixed
compensation
Corrected vol‐
ume flow
Fixed compensation
corrected volume flow
value that is used to
close the valve(s) be‐
fore the measured
amount value has
reached the target
amount
-
-
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
157
Содержание SITRANS F Coriolis FCT030
Страница 6: ...Table of contents FCT030 Modbus From firmware 4 0 6 Function Manual 06 2018 A5E39931813 AB ...
Страница 16: ...Communication 2 4 Connecting FCT030 Modbus From firmware 4 0 16 Function Manual 06 2018 A5E39931813 AB ...
Страница 194: ...Definitions A 1 Definitions FCT030 Modbus From firmware 4 0 194 Function Manual 06 2018 A5E39931813 AB ...
Страница 198: ...Index FCT030 Modbus From firmware 4 0 198 Function Manual 06 2018 A5E39931813 AB ...