Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting options
Access
level
9386
Float / 4
Offset adjust‐
ment
Command to trim the
lower range value of
the loop current.
Simulate a loop current
of 4.0 mA, measure the
current with an ampere
meter and write the
measured value to this
register.
-
0 - 25
Write on‐
ly
9388
Float / 4
Gain adjustment Command to trim the
upper range value of
the loop current.
Simulate a loop current
of 20.0 mA, measure
the current with an am‐
pere meter and write
the measured value to
this register.
-
0 - 25
Write on‐
ly
Channel 4 frequency output settings
Table 4-44
Channel 4 frequency output settings
Modbus
register
Data type /
Size in bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting options
Access
level
9404
Unsigned / 2
Process value
Select process value
for the current output.
0
● 0: Mass flow
● 1: Volume flow
● 2: Density
● 3: Medium
temperature
Read /
write
9410
Unsigned / 2
Flow direction
Flow direction filter ap‐
plicable for Process val‐
ues (9404) set to 0, 1,
19, or 27.
0
● 0: Positive direction
● 1: Negative direction
● 2: Positive and
negative directions
● 3: Positive and
negative directions
(symmetric mode)
Read /
write
9435
Float / 4
Frequency val‐
ue high
Upper frequency value.
Upper nominal output
range boundary.
10000.0
[Hz]
0.0 - 12500.0
Read /
write
9437
Float / 4
Frequency val‐
ue low
Lower frequency value.
Lower nominal output
range boundary.
0.0 Hz
0.0 - 12500.0
Read /
write
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
112
Function Manual, 06/2018, A5E39931813-AB
Содержание SITRANS F Coriolis FCT030
Страница 6: ...Table of contents FCT030 Modbus From firmware 4 0 6 Function Manual 06 2018 A5E39931813 AB ...
Страница 16: ...Communication 2 4 Connecting FCT030 Modbus From firmware 4 0 16 Function Manual 06 2018 A5E39931813 AB ...
Страница 194: ...Definitions A 1 Definitions FCT030 Modbus From firmware 4 0 194 Function Manual 06 2018 A5E39931813 AB ...
Страница 198: ...Index FCT030 Modbus From firmware 4 0 198 Function Manual 06 2018 A5E39931813 AB ...