Group
Bit
Alarm
ID
Restrictions
Associated alarm
class
Alarm items 2
(Sensor alarms 2)
0
Auxiliary input failure
32
1
DSL Functional Safety event
33
2
Unreliable flow measurement
34
3
Failure on channel 5 or channel 6
35
4
Configuration error 2
36
5
DSL system monitor
37
Alarm items 4
(Process alarms 1)
0
Mass flow above alarm limit
96
PA
1
Mass flow above warning limit
97
PW
2
Mass flow below warning limit
98
PW
3
Mass flow below alarm limit
99
PA
4
Volume flow above alarm limit
100
PA
5
Volume flow above warning limit
101
PW
6
Volume flow below warning limit
102
PW
7
Volume flow below alarm limit
103
PA
8
Density above alarm limit
104
PA
9
Density above warning limit
105
PW
10
Density below warning limit
106
PW
11
Density below alarm limit
107
PA
12
Medium temperature above alarm limit
108
PA
13
Medium temperature above warning limit 109
PW
14
Medium temperature below warning limit
110
PW
15
Medium temperature below alarm limit
111
PA
16 -
31
Reserved
MODBUS commands
4.10 Modbus holding registers tables
FCT030 Modbus (From firmware 4.0)
146
Function Manual, 06/2018, A5E39931813-AB
Содержание SITRANS F Coriolis FCT030
Страница 6: ...Table of contents FCT030 Modbus From firmware 4 0 6 Function Manual 06 2018 A5E39931813 AB ...
Страница 16: ...Communication 2 4 Connecting FCT030 Modbus From firmware 4 0 16 Function Manual 06 2018 A5E39931813 AB ...
Страница 194: ...Definitions A 1 Definitions FCT030 Modbus From firmware 4 0 194 Function Manual 06 2018 A5E39931813 AB ...
Страница 198: ...Index FCT030 Modbus From firmware 4 0 198 Function Manual 06 2018 A5E39931813 AB ...