The device provides means to remap 20 Modbus registers.
Table 4-3
Modbus register mapping
Modbus
register
Data type /
Size in
bytes
Parameter
Description
Default value
[units]
(units register)
Value range /
Setting op‐
tions
Access
level
10448
Unsigned /
4
Enable mapping
Activation/deactivation of
the register mapping. A set
bit means that the mapping
pair is activated, a cleared bit
that the mapping pair is de‐
activated.
Bit 0:
Requested register 1 / Tar‐
get register 1
…
Bit 19:
Requested register 20 / Tar‐
get register 20
0
0 - 1048575
Read /
write
10450
Unsigned /
2
Register 1 source
Modbus register that ap‐
pears within Modbus re‐
quest is redirected to the pa‐
rameter specified by Target
register 1
65535
0 - 65535
Read /
write
10451
Unsigned /
2
Register 1 target
Register of an existing prod‐
uct parameter to which a
Modbus request is redirected
65535
0 - 65535
Read /
write
...
10488
Unsigned /
2
Register 20 source Modbus register that ap‐
pears within Modbus re‐
quest is redirected to the pa‐
rameter specified by Target
register 20
65535
0 - 65535
Read /
write
10489
Unsigned /
2
Register 20 target
Register of an existing prod‐
uct parameter to which a
Modbus request is redirected
65535
0 - 65535
Read /
write
MODBUS commands
4.4 Modbus register mapping
FCT030 Modbus (From firmware 4.0)
Function Manual, 06/2018, A5E39931813-AB
37
Содержание SITRANS F Coriolis FCT030
Страница 6: ...Table of contents FCT030 Modbus From firmware 4 0 6 Function Manual 06 2018 A5E39931813 AB ...
Страница 16: ...Communication 2 4 Connecting FCT030 Modbus From firmware 4 0 16 Function Manual 06 2018 A5E39931813 AB ...
Страница 194: ...Definitions A 1 Definitions FCT030 Modbus From firmware 4 0 194 Function Manual 06 2018 A5E39931813 AB ...
Страница 198: ...Index FCT030 Modbus From firmware 4 0 198 Function Manual 06 2018 A5E39931813 AB ...