17. Open the “System function” menu under “Maintenance”.
18. Click “Warm start”.
7.7
Recognizing the safe operating state
7.7.1
Recognizing the safe operating state
Prerequisites
•
A leak tightness check was successfully performed.
Procedure
1.
Check the power supply is within the specification.
2.
Check clean instrument air is connected and available according to specification.
3.
Check the flow rate is approx. 100 l/h.
4.
Check all housing doors are closed.
5.
Check only the green LED lights on the display.
6.
Check all measured values are plausible.
COMMISSIONING
7
8021339/1CHC/V1-1/2021-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
61