6.
Set the “Enable” check mark to deactivate the measuring point or remove it to
activate the measuring point.
7.
Check “Live view” so that the settings are used by the BCU.
7.5.11
Setting analog outputs with SOPAS ET
Overview
The analog outputs are preset.
If necessary, the output ranges can be changed.
Prerequisites
•
SOPAS ET is connected to the measuring device.
Procedure
1.
Open the BCU module with a double-click.
2.
Open the “Analog outputs” menu via “Parameter” and “I/O”.
3.
Remove the check mark for “Live view”.
4.
Mark the corresponding analog output.
5.
Click “Edit”.
6.
Change the output ranges. The start value is always 4 mA and the end value 20
mA.
7.
Activate the required output areas with a check mark.
✓
Switching is automatically controlled with two active output ranges.
8.
Click “Save”.
7.5.12
Testing analog outputs with SOPAS ET
Overview
Testing the analog outputs checks whether they output correctly.
Prerequisites
•
SOPAS ET is connected to the measuring device.
Procedure
1.
Open the BCU module with a double-click.
2.
Open the “Analog outputs” menu via “Maintenance” and “Tests”.
3.
Mark the analog output to be tested.
4.
Click “Save”.
✓
The Test menu is displayed.
5.
Enter the configured end value as the test value.
6.
Check on the I/O module that the output is 20 mA.
7.
The measured mA value can be readjusted when it deviates from the defined test
value.
Additional information
Other concentrations can also be specified as test values. AO(n)O [mA] shows the
expected value output by the I/O module.
7.6
Performing the leak tightness check
Prerequisites
•
SOPAS ET is connected to the measuring device.
•
If there are several measuring points: select the corresponding measuring point
before the test.
COMMISSIONING