Logbook
text (Dis‐
play)
Logbook
text (SOPAS
ET)
Description
Interpretation and possibly
correction
X X F Heater 1
F Heater 1
Heater 1 sensor or power
control defective
Ambient temperature out‐
side specification or sensor
module is defective and
must be replaced.
X
F Heater 2
F Heater 2
Heater 2 sensor or power
control defective
Ambient temperature out‐
side specification or sensor
module is defective and
must be replaced.
X
F Heater 3
F Heater 3
Heater 3 sensor or power
control defective
Ambient temperature out‐
side specification or sensor
module is defective and
must be replaced.
X
F Heater 4
F Heater 4
Heater 4 sensor or power
control defective
Ambient temperature out‐
side specification or sensor
module is defective and
must be replaced.
X
F Heater 5
F Heater 5
Heater 5 sensor or power
control defective
Ambient temperature out‐
side specification or sensor
module is defective and
must be replaced.
X X X E SPI 1
E SPI 1
Malfunction of SPI 1 data
transmission
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
X X E SPI 2
E SPI 2
Malfunction of SPI 2 data
transmission
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
X
E I2C 1
E I2C 1
Malfunction of I2C 1 data
transmission
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
X X E I2C 2
E I2C 2
Malfunction of I2C 2 data
transmission
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
X X X E I2C 3
E I2C 3
Malfunction of I2C 3 data
transmission
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
X X X F EEPROM
F EEPROM
EEPROM malfunction
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
X X X E LM75
E LM75
Malfunction of LM75 (tem‐
perature measurement)
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
X X F Config.
invalid
F Config.
invalid
Error in configuration
Error when loading sensor
parameters, restart device
X X F Program
sequence
F Program
sequence
Error in program timing
Malfunction in program
sequence, no action possi‐
ble. Restart device. If error
is repeated, replace module
11
TROUBLESHOOTING
102
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
8021339/1CHC/V1-1/2021-07 | SICK
Subject to change without notice