WARNING
Health hazard due to toxic compounds in the condensate
b
Observe all safety regulations for the application.
b
In case of contact with the skin or eyes, rinse the affected parts immediately with
clear water and consult a doctor.
NOTE
The component is system-specific. The described procedure is therefore exemplary and
can deviate.
Prerequisites
•
Measuring device is switched off.
Procedure
1.
Note the date of installation on the new water trap.
2.
Loosen the cap nuts of the sample gas hoses on the water trap side.
3.
Pull the hoses out of the screw fitting.
4.
Insert a new water trap.
5.
Reattach the tube connections.
6.
Check the gas tightness of the sample gas path.
Related topics
•
Further information on attaching the tube connections:
see "Using a plastic fitting",
•
Further information on the leak test:
see "Performing the leak tightness check",
10.10.6
Replacing the filter pad
Important information
NOTICE
Dirt can get into the measuring device when replacing the filter pad.
b
Only replace the filter pad when the measuring device is switched off.
Prerequisites
•
Measuring device is switched off.
Procedure
1.
Open the cover of the fan.
2.
Take filter pad out.
3.
Insert new filter pad.
4.
Close the cover.
MAINTENANCE
10
8021339/1CHC/V1-1/2021-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
85
Subject to change without notice