Logbook
text (Dis‐
play)
Logbook
text (SOPAS
ET)
Description
Interpretation and possibly
correction
Sensor lost
F Sensor lost Error: Connection to FINOR
sensor lost
Check wiring to FINOR sen‐
sor. Restart device. Replace
complete analyzer module if
necessary
Sensor lost
F Sensor lost Error: Connection to Gas
module lost
Check wiring to gas mod‐
ule. Restart device. Replace
complete analyzer module if
necessary
Sensor
U Sensor
DEFOR sensor status
uncertain
Check status messages at
sensor and clear messages
Sensor
U Sensor
FINOR sensor status uncer‐
tain
Check status messages at
sensor and clear messages
Sensor
U Sensor
Gas module status uncer‐
tain
Check status messages at
sensor and clear messages
Sensor regis‐
tration error
Sensor regis‐
tration error
Error during DEFOR sensor
registration
Check sensor wiring. Restart
device.
Sensor regis‐
tration error
Sensor regis‐
tration error
Error during FINOR sensor
registration
Check sensor wiring. Restart
device.
Sensor regis‐
tration error
Sensor regis‐
tration error
Error during Gas module
sensor registration
Check sensor wiring. Restart
device.
Sensor error
A
Sensor error
A
Error in data communica‐
tion to DEFOR sensor
Check sensor wiring. Restart
device.
Sensor error
A
Sensor error
A
Error in data communica‐
tion to FINOR sensor
Check sensor wiring. Restart
device.
Sensor error
A
Sensor error
A
Error in data communica‐
tion to gas module
Check sensor wiring. Restart
device.
Sensor error
B
Sensor error
B
Error in data communica‐
tion to DEFOR sensor
Check sensor wiring. Restart
device.
Sensor error
B
Sensor error
B
Error in data communica‐
tion to FINOR sensor
Check sensor wiring. Restart
device.
Sensor error
B
Sensor error
B
Error in data communica‐
tion to gas module
Check sensor wiring. Restart
device.
Sensor de-
registration
error
Sensor de-
registration
error
Error during registration of
a module
Check sensor wiring. Restart
device.
Measuring
screen
Measuring
screen
Error in display parameters Error in BCU configuration
display parameters
Tag: Formula Tag: Formula Error in formula interpreter Error in BCU configuration in
formula part
Cyclic trigger Cyclic trigger Error in timer configuration Error in BCU during timer
configuration
Adj./Val.
Adj./Val.
Error in adjustment/valida‐
tion configuration
Error in BCU in adjust‐
ment/validation configura‐
tion
Factory set‐
tings backup
failed
Factory set‐
tings backup
failed
Backup of factory settings
failed
Error during data backup.
Repeat data backup
User set‐
tings backup
failed
User set‐
tings backup
failed
Backup of user settings
failed
Error during data backup.
Repeat data backup
TROUBLESHOOTING
11
8021339/1CHC/V1-1/2021-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
105
Subject to change without notice