7.5.2
Setting the heating controller
Overview
1
Actual temperature
2
Temperature setpoint
3
Arrow keys for setting
The heating controller regulates the temperature of the sample gas line and the gas
sampling unit.
The status signal is passed to the BCU of the analyzer when the heating controller
reaches the setpoint temperature.
Procedure
1.
Use the arrow keys on the heating controller to set the desired temperature. The
recommended temperature is 180 °C.
✓
If the arrow key is not pressed for 3 seconds, the set temperature is saved as the
setpoint temperature.
7.5.3
Setting the cooler temperature
Overview
1
Display
2
Cancel
3
Arrow up
4
Confirm
5
Arrow left
6
Arrow down
COMMISSIONING
7
8021339/1CHC/V1-1/2021-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
55