4.
Connect the 10 mm hose to the smaller connection on the lid of the condensate
container.
5.
Connect the other end to the reducer.
5.5.13
Connecting the filling level sensor to the condensate container (option)
Prerequisites
•
Condensate container and metal stand are attached.
Procedure
1.
Connect the cable to the level sensor.
2.
Fit the level sensor on the retaining plate using the screws supplied.
3.
Decide how often the condensate container has to be emptied and install the
appropriate level sensor.
°
1
At the top
of the container: Occasional emptying
°
2
At the bottom
of the container: Frequent emptying
1
Level sensor at top of the container
2
Level sensor at bottom of the container
4.
Insert the cable through the corresponding cable entry.
5.
Lay the cable in the cable duct, strip the insulation and connect it with a ferrule.
6.
Adjust the sensitivity of the level sensor with a screwdriver.
1
Adjustment screw
2
Status LED
°
Turn adjustment screw counterclockwise until end position is reached.
°
Turn clockwise until LED turns on.
°
Slowly turn counterclockwise until LED turns off.
°
From the point where the LED goes out, make an additional quarter turn
counterclockwise.
✓
The status LED indicates whether the sensor is triggered or not and sends an
MOUNTING
5
8021339/1CHC/V1-1/2021-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | MARSIC280