7.4.11
Setting automatic validation and adjustment with SOPAS ET
Overview
Automatic validation and adjustment is not set up by default.
Prerequisites
•
SOPAS ET is connected to the measuring device.
Procedure
1.
Open the BCU module with a double-click.
2.
Open the “Adjustment/validation” menu via “Parameter”.
3.
Remove the check mark for “Live view”.
4.
Select the appropriate component.
5.
Click “Edit”.
6.
Set up the validation/adjustment:
°
Start time = desired time (hh:mm)
°
Start date = desired date for the first start (yy-mm-dd)
°
Period = Repeat interval and frequency (Hour, Day, Week)
°
Active = Activation of automatic adjustment
7.
Click “Save”.
8.
Set the check mark for “Live view”.
7.5
Parameterization
7.5.1
Setting the time
Overview
Figure 17: BCU: Internal clock
Prerequisites
•
SOPAS ET is connected to the measuring device.
Procedure
1.
Open SOPAS ET.
2.
Connect the device
3.
Navigate to the Parameterization/Device menu.
4.
Click “PC-Time”.
✓
Date and time setting are synchronized to the PC.