Important information
NOTE
The purge time must be at least 50 seconds. Otherwise, stable measured values
cannot be guaranteed with several measuring points. The sample gas needs this time
to be stable in the analyzer during a sample change.
The total time of all samples together may not exceed 285 seconds.
Prerequisites
•
SOPAS ET is connected to the measuring device.
Procedure
1.
Open the BCU module with a double-click.
2.
Open menu “Measuring point automatic” under “Parameter”.
3.
Remove the check mark for “Live view”.
4.
Select the appropriate measuring point.
5.
Click “Edit”.
✓
The settings menu of the measuring point opens.
6.
The purging and measuring times can be adapted to the actual conditions.
7.
Set the check mark for “Live view” so that the settings are used by the BCU.
7.5.9
Activating and deactivating measuring points
Overview
If the measuring device has more than one measuring point, unused measuring points
can be deactivated if required.
Procedure
1.
Press “Menu”.
2.
Select menu “5 Parameter”.
3.
Select menu “10 Special”.
4.
Select menu “1 Meas. Pt (MP)”.
5.
Select the appropriate measuring point.
6.
Select menu “1 On/Off”.
7.
Selecting “On” activates the measuring point. Selecting “Off” deactivates the
measuring point.
8.
Press “Save” to confirm.
7.5.10
Activating and deactivating measuring points with SOPAS ET
Overview
If the measuring device has more than one measuring point, unused measuring points
can be deactivated if required.
Prerequisites
•
SOPAS ET is connected to the measuring device.
Procedure
1.
Open the BCU module with a double-click.
2.
Open menu “Measuring point automatic” under “Parameter”.
3.
Remove the check mark for “Live view”.
4.
Select the appropriate measuring point.
5.
Click “Edit”.
✓
The settings menu of the measuring point opens.