Logbook
text (Dis‐
play)
Logbook
text (SOPAS
ET)
Description
Interpretation and possibly
correction
X
F Motor 2
position
F Motor 2
position
Motor 2 position error
Restart device. If error is
repeated, replace complete
analyzer module
X
F Motor 3
position
F Motor 3
position
Motor 3 position error
Restart device. If error is
repeated, replace complete
analyzer module
X
M Extrane‐
ous light
M Extrane‐
ous light
Extraneous light has pene‐
trated the measuring sys‐
tem
Device open, extraneous
light has penetrated the
measuring system
X
M Mirror
M Mirror
Mirror error
Restart device. If error is
repeated, replace complete
analyzer module
X
M Beam
splitter
M Beam
splitter
Beam splitter error
Restart device. If error is
repeated, replace complete
analyzer module
X
M Filter
M Filter
Filter error
Restart device. If error is
repeated, replace complete
analyzer module
X X M UV source
intensity / M
IR intensity
M UV source
intensity / M
IR intensity
Source intensity below
maintenance request
threshold
Replace complete analyzer
module
X X X C mainte‐
nance mode
C Mainte‐
nance active
Maintenance active
Status message, check
active, maintenance active
X X X C Start test
C Start test
Start check
Function checks active after
switching on
X X X U mainte‐
nance mode
U Mainte‐
nance active
Maintenance active
Maintenance active, meas‐
ured values unreliable, no
action required
X X X U Start test
U Start test
Start check
Function checks active after
switching on, no action
required
X
U Extrane‐
ous light
U Extrane‐
ous light
Extraneous light has pene‐
trated the measuring sys‐
tem
Measured value unreliable,
device open, extraneous
light has penetrated the
measuring system
X
U Mirror
U Mirror
Mirror error
Restart device. If error is
repeated, replace complete
analyzer module
X
U Beam
splitter
U Beam
splitter
Beam splitter error
Restart device. If error is
repeated, replace complete
analyzer module
X
U Filter
U Filter
Filter error
Restart device. If error is
repeated, replace complete
analyzer module
X X U UV source
intensity / U
IR intensity
U UV source
intensity / U
IR intensity
Source intensity below
uncertain threshold
Restart device. If error is
repeated, replace complete
analyzer module
X
U position
motor 1
U position
motor 1
Motor 1 position error
Motor 1 has not found
the zero position. Restart
device. If error is repeated,
replace complete analyzer
module
11
TROUBLESHOOTING
100
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
8021339/1CHC/V1-1/2021-07 | SICK