Important information
WARNING
Health hazard due to toxic compounds in the condensate
b
Observe all safety regulations for the application.
b
In case of contact with the skin or eyes, rinse the affected parts immediately with
clear water and consult a doctor.
Procedure
1.
Loosen tube connections on the pump.
2.
Press the wrapping tape roll together at the grip recesses and turn the S-bolt
clockwise to the stop.
3.
Remove the wrapping tape roll and pull the old pump hose out of the guide using
the hose sleeves.
4.
Press the press rollers together and check whether spring tension is present, if
not, replace the complete hose pump.
5.
Insert the new pump hose with hose nozzles into the guides of the wrapping tape
roll.
6.
Insert the wrapping tape roll complete with hose into the dovetail guide of the
pump body.
7.
Press the wrapping tape roll together at the grip recesses and turn the S-bolt
counter-clockwise to the stop.
10.10.5
Replacing the water trap
Overview
Figure 24: Water trap (system-specific)
Important information
CAUTION
Poisoning hazard through sample gas
Sample gas may be present in the components in contact with the sample gas after
switching off.
b
Set the measuring device to Standby mode before turning it off.
10
MAINTENANCE
84
O P E R A T I N G I N S T R U C T I O N S | MARSIC280