10.10.7
Emptying the condensate container.
Important information
CAUTION
Risk of chemical burns by acid gas
Acid condensate could escape when working on the sample gas lines and the associ‐
ated assemblies.
b
Take appropriate protective measures for work (e.g., by wearing a safety mask,
protective gloves and acid resistant clothes)
b
In case of contact with the eyes, rinse immediately with clear water and consult a
doctor.
Procedure
1.
Empty the condensate container when the alarm signal is triggered.
✓
The alarm stops automatically after emptying the container.
10.10.8
Replacing the analyzer
10.10.8.1
Remove the analyzer
Overview
Figure 25: Sample gas connections on
analyzer
1
Sample gas inlet (seen from
the front, bottom left)
2
Sample gas outlet (seen from
the front, bottom right)
Figure 26: Electrical connections in analyzer
1
Power voltage connection at terminal
strip
2
Signal connections (I/O)
3
Distributor board with Ethernet for
MPR (option), Modbus, service inter‐
face (LAN)
10
MAINTENANCE
86