Important information
WARNING
The UV lamp emits UV radiation. Exposure to UV radiation may result in eye or skin
irritation.
b
Use suitable shielding during operation of the UV lamp.
Prerequisites
•
The measuring device is switched off and secured against being switched on
again.
Procedure
1.
Unscrew and fold open the housing door of the analyzer bottom part.
2.
Unscrew the grounding conductors on the side panel.
3.
Mark and unscrew the sample gas inlet and gas outlet at the connections of the
complete analyzer module.
4.
3 RJ45 plugs (CAN-Bus): Press latch and pull off.
5.
3 24 V plugs: Press latch and pull off.
6.
Unscrew 7 black screws from the base plate.
7.
Pull the complete analyzer module by the grip handles to the front and lift out
10.10.9.2
Fit the complete analyzer module
Overview
The complete analyzer module contains all measuring modules.
It is in the analyzer bottom part.
It can be removed complete with all modules.