Display
Function
Enter
Call up/start selected menu function
Menu
Call up the main menu.
If the
Menu>
button is not shown: First press
MEAS
.
Save
Save input/exit
Set
Start setting
Select
Select function/character
Start
Start procedure
In a selection list: Move cursor upwards
During input: Next character
Move cursor downwards
Move cursor to the left
Move cursor to the right
8.2
Status messages
8.2.1
Opening the logbook
Overview
The logbook displays the error messages present on the device. The Status column
shows whether a maintenance or error message is active.
Status “On” = active message
Status “Off” = message no longer pending
Important information
NOTE
The device must heat up for about two hours. Many error messages can be related to
an operating temperature that has not been reached.
Therefore the logbook should only be checked after the heating time.
Prerequisites
•
SOPAS ET is connected to the measuring device.
Procedure
1.
Select menu “Logbook” in the “Diagnosis” folder.
2.
Check the messages from the measuring modules (DEFOR, FINOR, Gas module)
before those from the BCU.
8.2.2
Checking the DEFOR hours of operation
Overview
The DEFOR lamp must be replaced regularly for maintenance reasons.
•
Replace the DESOX-DEFOR module after 4 years (35 000 hours)
Procedure
1.
Open the “Hours of operation” menu in the “Diagnosis” folder.
2.
Check the hours of operation of the DEFOR lamp and replace when necessary.
OPERATION
8
8021339/1CHC/V1-1/2021-07 | SICK
O P E R A T I N G I N S T R U C T I O N S | MARSIC280
63