07931001B revision 6
© Servomex Group Limited. 2020
109
Register Name
M1
Address
M2
Address
Read/
Write
Notes
set scaling correction
factor
50059
50139
R/W
set scaling correction factor for auto val on
measurement 1
Measurement
Concentration
50061
50141
R
The current value of validation concentration.
This can be read at any time during auto val
concentration standard
deviation
50063
50143
R
The current value of validation concentration
standard deviation. This can be read at any
time during auto val. The std dev is expressed
as a percentage of the current validation
concentration
transmission standard
deviation
50065
50145
R
The current value of validation transmission
standard deviation. This can be read at any
time during auto val. The std dev is expressed
as a percentage of the current transmission
zero measurement
scaled to cell length
50067
50147
The zero measurement calculated by the
Laser 3 Plus during auto validation
span measurement
scaled to cell length
50071
50151
The span measurement calculated by the
Laser 3 Plus during auto validation
Final zero threshold
50075
50155
The cell concentration measurement must be
below this level after the final zero flush. If it is
not, then a Maintenance Required fault will be
raised on exiting auto validation. This is to alert
the user that there may be span-gas in the
validation cell and that subsequent measurements
may be incorrect.
Password
50077
50157
Set to operator password to enable auto validation
The password must be entered before each run of
auto val
Modbus Transaction Frequency
Modbus is not a high-speed data interface. Experience with the Laser 3 Plus has
shown that a minimum 100mSec delay should be enforced between successive
Modbus transactions. If attempts are made to read or write Modbus registers any
faster than this it may result in bus errors.
Modbus Addresses That Should Be Monitored Continuously By
The HMI
The Laser 3 Plus does not have the ability to control the gas valves so this
responsibility has to be assumed by the HMI. The Laser 3 Plus uses a set of
Modbus addresses to tell the HMI when to turn the gases on and off.
The registers are
•
Requested zero-gas state
•
Requested span-gas state.
Refer to the Modbus map in section 12.1.14 for the register addresses.
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...