176
© Servomex Group Limited. 2020
07931001B revision 6
11. Appendix Display unit conversion
When you select display units as described in Section 4.8, you must ensure that you also
enter the correct unit conversion factor, as shown in the table below:
To convert
from *
to
†
use the units conversion factor applicable
gas(es)
%
ppm
10000
any
ppm
%
0.0001
any
ppm
vpm
1
any
ppm
mg/m3
1.4277
O2
"
"
0.750
NH3
"
"
1.2492
CO
"
"
1.9631
CO2
%
mg/m3
14277
O2
"
"
7500
NH3
"
"
12492
CO
"
"
19631
CO2
ppm
%LEL
0.0008
CO
%
%LEL
8
CO
%
mol/mol
0.01
any
ppm
mol/mol
#
#
* Measurement default units
†
Selected display units
#
This conversion is not supported
Note:
To return to the measurement default units, select the "off" units
selection option and set the units conversion factor to "1":
see Section 4.8
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...