Viewing Validation Results
The Laser 3+ analyzer records the results of all validation operations. The validation
results can be viewed as follows:
Step No
Display
Description
1
Select
“
Calibrate
”
from the top-level menu.
2
Select “View History”.
3
You can then see all of the recorded results
for both calibration and validation. Use the up
and down arrows to scroll between records.
The right arrow can be used to see more
parameters within each record.
T = Span Target concentration
R = Span concentration Result