07931001B revision 6
© Servomex Group Limited. 2020
57
Figure 4-18: Select alarm mode menu
Latching and non-latching alarms
You can configure the two measurement alarms to be either latching or non-
latching:
Alarm setting
Meaning
Latching
Once an alarm is raised, it remains activated until the alarm is manually
unlatched
Non-Latching
Once an alarm is raised, it remains activated only until a subsequent
‘good’ sample measurement is made (i.e. one that would not trigger an
alarm). The alarm condition is then reset
Alarms
Alarm 1 : 2 Trans
Alarm 1 : 2 Trans
Alarm 1 : 2 Trans
Active
Alarm
Measurement
Mode None
Set Up
Low
Follow
1 (Tx 2)
2Trans %
High
History
Select
mode
Alarm 1 : 2 Trans
Alarm 1 : 2 Trans
Alarm 1 : 2 Trans
Hysteresis
Level
Latching
Yes
0.00%
0.00%
No